CVE-2024-48966
Essential information
- Published
- 14/11/2024 22:15
- Modified
- 15/11/2024 13:58
- Author
- —
- Creator
- —
- CVSS
- 10.0 CRITICAL (v3.1)
- CISA KEV
- No
- CWE
- —
- CVSS vector
-
—
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H—
CVSS metrics
- Access vector
- —
- Access complexity
- —
- Authentication
- —
- Confidentiality impact
- —
- Integrity impact
- —
- Availability impact
- —
- Exploitability
- —
- Remediation level
- —
- Report confidence
- —
- Temporal score
- —
- Attack vector
- NETWORK
- Attack complexity
- LOW
- Privileges required
- NONE
- User interaction
- NONE
- Scope
- CHANGED
- Confidentiality impact
- HIGH
- Integrity impact
- HIGH
- Availability impact
- HIGH
- Exploit code maturity
- —
- Remediation level
- —
- Report confidence
- —
- Temporal score
- —
- Attack vector
- —
- Attack complexity
- —
- Attack requirements
- —
- Privileges required
- —
- User interaction
- —
- Confidentiality (V)
- —
- Confidentiality (S)
- —
- Integrity (V)
- —
- Integrity (S)
- —
- Availability (V)
- —
- Availability (S)
- —
- Exploit maturity
- —
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
NVD status
- Status
- Awaiting Analysis — CVE has been recently published to the CVE List and has been received by the NVD.
- Source
- [email protected]
- NVD
- View on NVD