CVE-2025-14095
Essential information
- Published
- 17/12/2025 12:15
- Modified
- 18/12/2025 15:08
- Author
- —
- Creator
- —
- CVSS
- 6.8 MEDIUM (v3.1)
- CISA KEV
- No
- CWE
- —
- CVSS vector
-
—
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H—
CVSS metrics
- Access vector
- —
- Access complexity
- —
- Authentication
- —
- Confidentiality impact
- —
- Integrity impact
- —
- Availability impact
- —
- Exploitability
- —
- Remediation level
- —
- Report confidence
- —
- Temporal score
- —
- Attack vector
- PHYSICAL
- Attack complexity
- LOW
- Privileges required
- NONE
- User interaction
- NONE
- Scope
- UNCHANGED
- Confidentiality impact
- HIGH
- Integrity impact
- HIGH
- Availability impact
- HIGH
- Exploit code maturity
- —
- Remediation level
- —
- Report confidence
- —
- Temporal score
- —
- Attack vector
- —
- Attack complexity
- —
- Attack requirements
- —
- Privileges required
- —
- User interaction
- —
- Confidentiality (V)
- —
- Confidentiality (S)
- —
- Integrity (V)
- —
- Integrity (S)
- —
- Availability (V)
- —
- Availability (S)
- —
- Exploit maturity
- —
Description
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.
Other related CVE's are CVE-2025-14096 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required configuration for Exposure:
Physical access to the analyzer is needed.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note:
CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
NVD status
- Status
- Awaiting Analysis — CVE has been marked for Analysis. Normally once in this state the CVE will be analyzed by NVD staff within 24 hours.
- Source
- 46b595e9-1acc-41cb-9398-adaf98d37a9b
- NVD
- View on NVD