CVE-2025-14096

· Published 17/12/2025 13:15 · Modified 18/12/2025 15:08

Labels: CVE-2025-14096 2025-12-1746b595e9-1acc-41cb-9398-adaf98d37a9bCVE-2025-14096

Essential information

Published
17/12/2025 13:15
Modified
18/12/2025 15:08
Author
Creator
CVSS
8.4 HIGH (v3.1)
CISA KEV
No
CWE
CVSS vector
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

CVSS metrics

Description

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.